HOW TO USE THE X-PCD
CLICK THE PLAY BUTTON ON THE VIDEO ABOVE
1. Open the end of the X-PCD that does not have holes (never open the end with holes)
2. Fold the X-Class II indicator from X-Labs in half (folding with bars inside or outside both work)
3. Place the indicator inside the X-PCD "legs' first - meaning the two legs of the indicator facing the bottom as this allows for easy removal
4. Close the lid tight
5. Place X-PCD anywhere in the chamber - any level in any position
6. When he cycle completes - remove the X-PCD (take precaution because it will be hot)!
7. Remove the X-Class II indicator
8. Check the X-Class II results match the validation sheet (all black)
9. Break the indicator in half, keeping only the 4 black bar results
10. Remove the sticker at the back of the X-Class II indicator and place into the logbook for your records